• Overview
  • Details
  • More
Share Products Categories Channel Cart

1/6
Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun图1Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun图2Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun图3Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun图4Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun图5Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun图6

Disposable IBP Transducer Kits Disposable Blood Pressure Sensor for B. Braun

Price $6.00
Min Order 40 Pieces  
Shipping from Guangdong, China
Quantity
-+
Product details
JIBPT-01-BL
Surgical Supplies Materials
Ethylene Oxide Sterilization
3 Years
Patient
With Logo Printing
Medical Materials & Accessories
Custom
CE/ISO13485
JCR
Blister Box
224*113*45mm
China
9018909991
Product Description

Product Description

Properties
Medical Materials & Accessories
Product Name
B.Braun Disposable IBP Transducer
Brand Name
JCR
Model Number
  JIBPT-01-BL

Detailed Photos

Main technical parameters and working environment requirements:
Disposable Pressure Transducer
Barometric pressure
70~106Kpa
Relative humidity
10~90% (non-condensing)
Operating pressure range
-50 ~ + 300 mmHg
Sensitivity
5.0μV/V/mmHg ±3%
Nonlinearity and hysteresis
full-scale reading±1.5%
Input impedance
1200Ω ~3200Ω
Zero pressure offset
-20mmHg~+20mmHg
Thermal offset shift
≤±0.3mmHg/°C
Offset drift
After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
Thermal span shift
≤±0.1%/°C
Frequency response
Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz
Defibrillator withstand
Defibrillation depends on the final connection of the equipment.
Leakage current
Leakage current is dependent on the final connection of the equipment
Overpressure load
-400~+4000mmHg
Shock resistance
Withstand falling from one meter for three times
 
Light sensitivity
When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage.
Storage and transport environment requirements
- 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.
Contact with the human body
≤168h
Intra-system velocity
Flow
2-5mL/h
Intra-system refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.
Product shelf-life
3 years


Company Profile

 

Exhibition

 

Certifications

Packaging & Shipping



Buyer review
Loading review details...
Contact

Shenzhen JCR Medical Technology Limited Company

Audited Supplier4 Years
ProfileCertified by SGS/BV

Message Type

* Content

* Name

* Phone

* Email

* Captcha

captcha

You May Like
Home > Products > Health & Medicine > Medical Supply > Disposable Medical Supplies