600II
Non-Customized
CE, ISO13485
Small
Infusion Equipment
Rate, Time, Body Weight , Sequential Mode
0.1~1500ml/H
≤ ±5%
3.0'' LCD Monochrome Screen, 240 X 128 Pixels
200 Drugs
1500 Events
8 Types,English, Spanish, French, Russian, Turkish
>4 Hours@25ml/H
IP34
Yes
Mindray
Seaworthy and Airworthy Packing
146 x 193 x 95mm
Shenzhen
9018
Product Description
Mindray Medical Pumps, Worthy of Trust.

Features:
Friendliness
Safety
Connectivity
Nurse call/ RS 232/ DC input


Specifications and parameters:

Features:
Friendliness
- Compact design with light weight
- 3" large LCD screen allows infusion status to be viewed easily
- Embedded handle allows for easy transportation
- Straightforward IV set installation
- Intuitive programming procedure with quick guidance on screen
Safety
- High accuracy: ±5%, satisfying neonatal use
- Water-proof membrane on fingers preventing leakage into pump body
- Anti-bolus, Drug name display, History record, Last therapy parameter memory
- IP34 dust and water protection
Connectivity
Nurse call/ RS 232/ DC input


Specifications and parameters:
| Dimension | 146 x 193 x 95mm |
| Infusion set | Applicable to GB 8368-2005 disposals; diametersΦ3.5-4.5mm, thickness 0.8-1.2mm |
| Infusion mode | Rate mode, Time mode, Body weight mode, Sequential mode |
| Flow rate range | 0.1~1500ml/h |
| Accumulated volume | 0.1~9999 ml |
| VTBI | 0.1~999.9 ml, increment is 0.1 ml; 1000-9999ml, increment is 1ml |
| Increment | 0.1~999.9 ml/h, increment is 0.1 ml/h; 1000~1500 ml/h, increment is 1 ml/h |
| Accuracy | ≤ ±5% |
| KVO rate | 0.1-5.0 ml/h, adjustable, increment is 0.1ml/h |
| Bolus rate | 0.10-1500ml/h |
| Purge rate | 800ml/h |
| Occlusion | Low 150±125 mmHg (20kPa); Medium 525±125 mmHg (70kPa); High 900±180 mmHg (120 kPa) |
| Air bubble detection | Single bubble detection: 50, 100, 250, 500, 800l μl Accumulated bubble detection: 0.1-4.0ml/h, adjustable |
| Screen | 3.0'' LCD monochrome screen, 240 x 128 pixels |
| Titration | Yes, increment is 0.1ml/h |
| Keypad lock | Yes |
| Pause | Yes |
| Self-test system | Yes |
| Last therapy parameter memory | Yes |
| Drug library | 200 drugs |
| History record | 1500 events |
| Weight | <1.6kg (not including clamp) |
| Handle | Embedded handle |
| Mounting type | Standard: Vertical Pole |
| Alarm Volume | 1-8 levels selectable |
| Language | 8, Chinese, English, Spanish, French, Russian, Turkish, Portuguese, German |
| Alarm | VTBI done, VTBI near done, KVO finish, Bottle empty, door open, IV set not installed, occlusion, air in line, drop error, system error, system abnormal, batter y low, batter y empty, reminder, AC power disconnection, standby end |
| Safety Requirement | |
| AC power | 100-240V (50/60Hz, 0.8-0.28A) |
| DC power | 10-16V (4.0-2.5A) |
| Battery type | Lithium |
| Battery operation time | >4 hours@25ml/h |
| Charging time | ≤6 hours@100% |
| Classification | CF, Class I |
| IP Grade | IP 34 |
| Temperature | 5~40ºC for operating, -20~60ºC for storage |
| Humidity | 15~95% for operating, 10~95% for storage |
| Air pressure | 57~106kPa for operating, 50~106kPa for storage |
| Interface | |
| Data interface | Supported, RS232 |
| Nurse Call | Supported |
| DC input | Supported |