RD-850X
Customized
CE, FDA, EEC, ISO13485
Medium
Rd-850X
17"*17"
3072*3072
50kw
X Ray Machine
40kv-125kv
10mA ~ 630mA
100-240V, 50Hz/60Hz
125kv
25kHz
Dawei
Standard Export Package
3.5 M3
China
Product Description
Factory Price Hospital Radiography High Frequency Floor-Mounted Digital X-ray Equipment
| No. | Component | Performance parameter | |
| 1 | Medical diagnostic X-ray high-frequency high-voltage generator | Output Power: 50kW | |
| Input Voltage: 220VAC,±10%,50Hz | |||
| Inverter Frequency: 25kHz,±10% | |||
| kV range: 40 kV~150kV | |||
| mA range: 10 mA~630 mA | |||
| ms range: 1ms~6300 ms | |||
| mAs range: 0.1mAs~200mAs | |||
| Tech mode: kV-mA(AEC);kV-mAs;kV-mA-ms | |||
| APR: 2000 | |||
| AEC(Optional): AEC Ionization chamber power supply:+/-15V Minimum exposure time:5ms | |||
| Auxiliary power supply: +24V DC 2A;24V AC 8A | |||
| Communication Interface: RS-232 | |||
| 2 | X-ray Tube Assembly | X-ray Tube Voltage Range: 50kV~150kV | |
| Nominal Focal Spot Value: 0.6mm(Small Focus)/1.2mm(Large Focus) | |||
| Nominal Anode Input Power: 20kW(Small Focus)/50kW(Large Focus) | |||
| Anode Heat Content: 300KHu | |||
| 3 | Flat Panel Detector | Image Sensor: a-Si (Amorphous Silicon) TFT | |
| Scintillator: CsI:Tl | |||
| Effective Array: 3072 ×3072 | |||
| Pixel Size: 139 um | |||
| Effective Area (H x V): 427mm×427mm | |||
| Preview Time: 3s | |||
| Imaging Time: 5s | |||
| Spatial Resolution: 3.6 lp/mm | |||
| AD Conversion: 16 bit | |||
| Data Interface: Gigabit Ethernet | |||
| Trigger Mode: AED(optional)/Software/Prep | |||
| Dimension (L × W × H): 460mm×460mm×15mm | |||
| Weight: 4kg | |||
| Power Consumption: 20W | |||
| 4 | Collimator | Control Mode: Manual control | |
| Maximum Radiation: 150kV | |||
| Inherent Filtration Al. Equivalent: 1mm AL | |||
| Rectangular X-ray Field Selection Set Focus Distance, SID=100cm: Min 2cm×2cm;Max 43cm×43cm | |||
| X-Ray Field Deviation: ≤1% of SID | |||
| Light Field Indicator @ SID=100cm: ≥180lux | |||
| Light Field Indicator Precision: ≤1% of SID | |||
| Light source ON time: 30s | |||
| Maximum Radiation Leakage EN 60601-1-3: <0.5mGyh | |||
| Power Supply: 24V AC/DC,Max.4A | |||
| Dimension: 206mm×230mm×143mm | |||
| Weight: 7.0kg±0.5kg | |||
| 5 | Machine | Wall Stand | Motion control: Manual control |
| Overall height: 2125mm±10mm | |||
| BUCKY itinerary: 1450mm±10mm(The highest: 1625mm, the lowest: 335mm) | |||
| Starting force: ≤50N | |||
| Braking force: ≥100N | |||
| Tube Stand | Motion control: Manual control | ||
| Overall height: 2125mm±10mm | |||
| Longitudinal stroke of tube column: 1280mm±10mm | |||
| Tube rotation range: ±180° | |||
| SID:1120mm±10mm | |||
| Starting force: ≤50N | |||
| Braking force: ≥100N | |||
| Radiographic Table | Radiographic table height: 540mm±5mm | ||
| Table size (L×W): (2000mm±5mm)×(665mm±5mm) | |||
| BUCKY itinerary: Manual control, 1245mm±10mm | |||
| Table load-bearing: ≥200kg | |||
| 6 | Grid | Grid 1 | Grid density: 103 L/inch |
| Focusing distance: 100cm | |||
| Grid ratio: 10:1 | |||
| Grid 2 | Grid density: 103 L/inch | ||
| Focuing distance: 150cm | |||
| Grid ratio: 12:1 | |||
| 7 | Workstation | / | |
| 8 | Dawei DR image acquisition and processing software | This software is comprised of the following modules which provide a workflow of patient study: Patient Management: including patient registration, work list, and study management. Study operation: including body part selection, study items selection, and image acquiring. Image preview: including display, layout, and processing of image. Also tool options for advanced operation. Configuration: including configuration of system, study, and user management. Especially the configuration for worklist and storage. The image processing tools are as follows: Flip vertical, flip horizontal, rotate -90°, rotate 90°, free rotate, window/level change, auto WW/WL, ROI window/level, invert color, image stitching, reject image, zoom in/out, image advanced process, reset the image, invalid image, measure the angle, distance measure, point gray value of the image, ROI magnifier, magnifier, image hint, CTR(cardio-thoracic ratio) tool, arrow tool, copy tool, paste tool, email tool, cobb angle, monitor calibration tool, save as tool, compare tool and inverse compare tool. Other tools: mark left, mark right, mark anterior, text edit, clip tool, clip options tool, ellipse clip tool and delete tool. | |





Doctor recommendation